OPTI CCA Blood Gas Analyzer Machine , OPTI Medical
Please sign in so that we can notify you about a reply
Description
The OPTI CCA Blood Gas and Electrolyte Analyzer is a microprocessor-based analyzer that uses discrete sensors called optical electrodes to monitor optical fluorescence (optodes). OPTI CCA Blood Gas and Electrolyte Analyzer required for calibration, sample measurement, and waste containment is contained in a disposable, single-use cassette. Swiping the cassette package through the bar-code reader reads specific calibration information from the cassette into the analyzer. After that, the cassette is inserted into the measuring chamber.
By running a precise calibration gas mixture through the optode sensors, the OPTI CCA Blood Gas Analyzer heats the cassette to 37.0°C (0.1°C)/98.6°F (0.1°F) and performs a calibration verification on the sensors for PCO2 and PO2. Precision buffer solution in the cassette is used to calibrate the pH and electrolyte channels. The analyzer aspirates the blood sample into the cassette and across the optode sensors when the calibration is validated. Fluorescence emission is monitored after equilibration with the blood sample.
Calibration/Measurement
Sample Volume | 125 µL (60µL for B60 cassette) |
Calibration Time | <90 seconds |
Measurement Time | <120 seconds |
Typical Cycle Time | <210 seconds |
Operating Temperature | 10–30°C (50–86°F) |
Sample Type | Whole blood, serum, or plasma |
Measurement Principle | Optical fluorescence, reflectance |
Sample Application | Syringe, capillary or ComfortSampler |
Data Storage |
|
Dimensions/Weight
Height x Width x Depth | 124 x 362 x 230 mm (4.7 x 14.2 x 9.1 in) |
Weight | 4.3 kg (9.6 lb) with rechargeable battery pack |
Miscellaneous
Voltage/Frequency | 120 or 240 AC, 50–60 Hz |
Power Consumption (max.) | 110 VA |
Interface | ASTM, ASCII, POCT protocols via USB or Ethernet |
Approvals | CE, UL, IVDD, and FDA |
OPTI CCA-TS2 analyzer specifications:
Measured Parameters:
- pH, pCO2, pO2,
- Na, K, Cl Ca,
- THb, SaO2
- GLUCOSE, BUN, LACTATE
Salient Features:
- Storage of reagents at normal room temperature except BUN (2-8°C) and Lactate
- Automatic sample aspiration
- Portable (lightweight and Mains cum battery operated with 8 Hrs back-up)
- Electronic Quality control
- Barcode scanner for quality control information and can be conveniently entered using the on-board bar code scanner
- On-Board Thermal printer
- Display of results on the in-built active color touch screen
- 200 patient data
- RS 232 interface Infrared. ASCII
- CE, UL, IVDD Compliant
- FDA approved
- Up Gradable
Barometric Pressure: 300–800 mmHg
Operating Parameters:
- Minimum sample size: 125 ul
- Sample type: whole blood, serum, plasma
- Sample application: Syringe, capillary, or Micro sampler
- Sample Input: Automatic Aspiration
- Analysis time (time to result) < 120 sec.
- Ambient temp: 10°C – 32 °C
- Relative Humidity: 5% - 95% (non-condensing)
Calculated Parameters:
- Actual bicarbonate (HCO3), Base Excess (BE)
- Base Excess extracellular fluid, Base Excess Actual, Buffer basis (BB), Total CO2: (tCO2)
- Standard Bicarbonate, Standard pH(stpH)
- Oxygen content (O2ct), Hematocrit (Hct)
- Hydrogen ion concentration, Alveolar-Arterial: Oxygen difference (AaDO2),Anion gap(AG)
- Standardized ionized Calcium (pH = 7.4), p50
Input Parameters:
- Patient temp, Hemoglobin type ,p50, FlO2,RQ
- Operation ID. Patient ID No., Patient Sex
- Mean Corpuscular Hemoglobin Concentration (MCHC),Sample type art/ven/mix, ven/cap/cord/CPB
- User-defined field,3 User-defined fields:
- Puncture site: left radial, right radial, left brachial, right brachial, left femoral, right femoral, cord, scalp
Data Management:
- Print out: Built-in Thermal printer
- Interface: RS 232 C, ASCII or ASTM format
- Quality Control Data & standard reference cassette report: 1 month of 3 levels stored onboard.
Electrical & Gas supply :
- Voltage / Frequency: 100 – 250 VAC / 16 VDC, 2.8 A / 50 – 60 Hz.
- Power Consumption (Maximum): 110 VA
- Battery operation: Up to 8 hrs. (Rechargeable battery pack)
- Calibration Gas: Disposable small cylinder
Features
- From Manufatcurer
- On site
- 1 Years
Product Specifications
Brand/Manufacturer
Weight & Dimensions
Reviews