Cart is empty
Hygeia Digital 6 Channel Electrocardiograph 6T
Please sign in so that we can notify you about a reply
Description
TECHNICAL SPECIFICATIONS
Lead acquisition | 12bit/1000Hz (synchronously 12 leads) |
Work mode | Manual/Auto/1-channel 60s compressed record/Rhythm analysis |
Filter | Muscle filter: (25Hz/35Hz/45Hz/75Hz/100Hz,-3dB) AC filter: (50Hz or 60Hz, -20dB) Drift filter: (0.5Hz,-3dB) |
CMRR | ≥100dB (with AC Filter) |
Input circuit | Floating; Protection circuit against Defibrillator effect |
Input impedance | ≥50MΩ |
Input circuit current | ≤0.1µA |
Patient current leakage | ≤10µA |
Calibrating voltage | 1mV±2% |
Voltage tolerance | ≥±500mV |
Time constant | ≥3.2s |
Frequency response | 0.05~150Hz (-3db) |
Noise level | < 15µVp-p |
Inter-channel interference | ≤0.5mm |
Sensitivity | 1.25, 2.5, 5, 10, 20, 40mm/mV |
Recording mode | 3ch, 3+1ch, 6ch, 6+1ch |
Paper speed | 5, 6.25, 10, 12.5, 25, 50mm/s (±3%) |
Paper | 110mm x 20m roll paper |
Communication | RS232 / USB port |
Power supply | AC 1 10-230V (±10%), 50/60HZ (±1 Hz), 60VA DC Built-in rechargeable battery, 12.6V (1500mAh) |
Weight | 2.25Kg |
Box Dimension | 410mm x 320mm x 240mm |
Features
Estimated Delivery:
5 to 10 daysDays
Warranty :
- 1 Years
Product Specifications
CMMR:
105
dB
ECG Channel Print:
- 6 Channel
ECG Filter:
- AC Filter
- Drift Filter
Input Impedance :
50
MΩ
Printer /Recorder:
- Inbuilt Thermal Printer
Printing Paper Type & Size:
110mm
Recording Mode:
- Auto Mode
- Manual
- Rhythm
Brand/Manufacturer
Brand:
Hygia
Weight & Dimensions
Product Dimension (mm):
410mm x 320mm x 240mm
Product Weight:
2.25Kg
Power Supply & Battery
Battery capacity:
1500mAH