Contec CMS5100 Patient Monitor NIBP/SPO2/PR
Please sign in so that we can notify you about a reply
Description
MANUFACTURE: CONTEC MEDICAL SYSTEMS CO.,LTD
Model: CMS5100
CMS5100 Patient Monitor adopts Oscillometry for NIBP measuring, Photoelectric Oxyhemoglobin Inspection Technology combining Capacity Pulse Scanning & Recording Technology for SpO2 measuring. Systolic Pressure(SYS), Diastolic Pressure(DIA), Mean Pressure(MAP), SpO2 and Pulse rate(PR) can be accurately measured.
Features :
- Be applicable for adult, pediatric and neonatal for all-round monitoring, easy operation and high cost performance.
- Be applicable for medicine, surgery, operating room, ICU/CCU, emergency room, obstetrics and gynecology, pediatrics.
- Built-in rechargeable Li-polymer battery for uninterrupted monitoring.
- Compact and flexible appearance, easy for carrying and be suitable for indoor and outdoor(in ambulance) monitoring.
- With user-friendly interface.
- Display with 2.8'' (320×240) color TFT LCD and red and yellow highlight LED.
- Visual and audible alarm for Systolic Pressure(SYS), Diastolic Pressure(DIA), Mean Pressure(MAP), SpO2 and Pulse rate(PR), and upper and lower limit of alarm can be set.
- Unique Flash memory, storage for up to 2000 NIBP data and 30000 SpO2 data.
- Convenient and quick in inquiring measurement data, review for NIBP trend graph within 24-hour and SpO2 and Pulse rate(PR) trend graph within 20-hour.
Performance :
NIBP monitoring
Measurement method: Oscillometry
Measurement mode: manual/auto/continuous.
Auto measurement interval: adjustable from 1 to 90 minutes.
Resolution: 1mmHg.
Accuracy: Maximum Mean deviation≤±5mmHg, Maximum Standard deviation≤8mmHg.
Self-check function.
Overpressure protection: double protection for software and hardware.
Range :
Adult:
Systolic pressure: 40mmHg~270mmHg
Mean pressure: 20mmHg~235mmHg
Diastolic Pressure: 10mmHg~215mmHg
Pediatric:
Systolic pressure: 40mmHg~200mmHg
Mean pressure: 20mmHg~165mmHg
Diastolic Pressure: 10mmHg~150mmHg
Neonatal:
Systolic pressure: 40mmHg~135mmHg
Mean pressure: 20mmHg~110mmHg
Diastolic Pressure: 10mmHg~100mmHg
Upper and lower limit of alarm
Adult:
Systolic pressure: 40mmHg~270mmHg
Mean pressure: 20mmHg~235mmHg
Diastolic Pressure: 10mmHg~215mmHg
Pediatric:
Systolic pressure: 40mmHg~200mmHg
Mean pressure: 20mmHg~165mmHg
Diastolic Pressure: 10mmHg~150mmHg
Neonatal:
Systolic pressure: 40mmHg~135mmHg
Mean pressure: 20mmHg~110mmHg
Diastolic Pressure: 10mmHg~100mmHg
SpO2 monitoring
Continuous real-time measuring.
Adopt digital SpO2 technology, which has strong anti-interference capability.
Real-time display SpO2, pulse rate, SpO2 Plethysmogram, bar graph, SpO2 and Pulse rate(PR) trend graph.
Standard finger-clip probe for adult , binding probe for neonatal is optional.
SpO2 Measuring
Range: 35%~100% (Resolution: 1%)
Accuracy: ±2% in stage of 70%-100%, and unspecified in stage of less than 70%.
Pulse Rate Measuring
Range: 30bpm~250bpm(Resolution: 1bpm)
Accuracy: ±2bpm or ±2%, whichever is greater.
Measurement Performance in Weak Filling Condition: SpO2 and pulse rate can be shown
correctly when pulse-filling ratio is 0.4%, SpO2 error is ±4%; the Pulse rate error is ±2bpm or ±2%, whichever is greater.
Resistance to surrounding light: The deviation between the value measured in the condition of man-made light or indoor natural light and that of darkroom is less than ±1%.
Accessories :
A power cord
A power adapter
A SpO2 probe
An adult NIBP cuff
An extending tube for blood pressure
Physical characteristic :
Dimension for product: 190(L)×162(W)×240(H)mm
Weight: 1.6kg
Dimension for packaging: 402(L)×228(W)×320(H)mm
Features