Controls
- Conventional ventilation: continuous flow, pressure-limited, time-cycled
- Triggered ventilation: SIMV, SIPPV, PSV1), leak adapted
- Trigger: Flow/volume trigger, leak adapted
- Trigger delay: approx. 40-60 ms
- High Frequency Ventilation: CPAP HFV, IMV HFV; Frequency: 5 to 20 Hz
- Volume Guarantee Ventilation: SIMV VG, SIPPV VG, PSV VG
- Oxygen mixer loss (bleed flow): 0 (zero) L/min
Settings
- Inspiratory oxygen concentration: 21-100 vol.% oxygen
- Peak inspiratory pressure: 10 to 80 mbar
- PEEP/CPAP: 0 to 25 mbar
- Maximal frequency: 200 bpm
- Inspiratory time: 0.1 to 2 seconds
- Expiratory time: 0.2 to 30 seconds
- Inspiratory flow: 1 to 30 L/min
- Base Flow (VIVE): 1 to 30 L/min
Monitoring
- Flow monitoring: at the Y-piece, integrated
- Volume monitoring: at the Y-piece, integrated
- Lung function monitoring: compliance, resistance C20/C
- Method: linear regression analysis
- FiO2 monitoring: integrated
- Real-time curves: Flow and pressure, integrated
- Inspiratory oxygen concentration: 21-100 vol.% oxygen
- Peak pressure: to 99 mbar
- Mean airway pressure: to 99 mbar
- Graphic trends: 6 parameters, integrated
- Logbook: record of up to 100 alarms