Warranty :- 1 year
Product Condition :- Brand New
Delivery Time :- 5 to 10 days
Technical Secification
Lead
|
Standard 12 leads
|
lead acquisition
|
12bit/1000Hz (synchronously 12 leads)
|
Work mode
|
Manual / Auto / Physical Examination / Storage
|
Filter
|
- AC Filter :50Hz/60Hz
- EMG Filter: 25Hz / 45Hz
- Anti-Drift Filter: 0.15Hz (Adaptive)
|
CMMR
|
≥100dB (with AC Filter)
|
Input circuit
|
Floating; Protection circuit against Defibrillator effect
|
Input impedance
|
≥50MΩ
|
Input circuit current
|
≤0.05µA
|
Patient current leakage
|
≤10µA
|
Calibrating voltage
|
1 mV±2%
|
Voltage tolerance
|
≥±500mV
|
Time constant
|
≥3.2s
|
Frequency response
|
0.05~160Hz (-3db)
|
Noise level
|
< 15µVp-p
|
Inter-channel interference
|
≤0.5mm
|
Sensitivity
|
Auto, 2.5, 5, 10, 20, 40mm/mV
|
Recording mode
|
. 3ch+++, 6ch, 6ch+, 12ch
- Default Man. Mode is 3ch++ format
- Default Auto.Mode is 12ch format
|
Paper speed
|
6.25, 12.5, 25, 50mm/s (±3%)
|
Paper
|
215mm x 20m roll paper or 210mm x 140mm Z-fold paper
|
Communication
|
RS232 / USB port
|
Power supply
|
- AC 110-230V (±10%), 50/60HZ (±1 Hz), 60VA
- DC Built-in rechargeable battery, 14.4V (4400mmAh)
|
Dimension / Weight
|
347mm x 293mm x 83mm / 4.8Kg (ECG-3312B)
|
|