Medical UPS (isolated output)
specifications and product information

catalog number: 329-T035H 329-T070H 329-T120H
output capacity: 350 VA / 225 watts 700 VA / 450 watts 1000 VA / 750 watts
run time -half load: 46 minutes 42 minutes 34 minutes
run time -full load: 18 minutes 18 minutes 16 minutes
output regulation: 120 volts ±5% (running on battery power)
frequency regulation: 60Hz ± 0.5Hz on battery power
output waveform: pulse width modulated sine wave
input voltage rating: 120 volts 60Hz nominal
leakage current: less than 300 ua including connected loads
maximum input current: 2.3 amps 7.8 amps 9.2 amps
line voltage regulation: 115 volts –13% / +20% 120 volts –14% / +8% 120 volts –14% / +8%
transfer time: 2-4 ms 2-4 ms 2-4 ms
transfer to battery: 83 volts,   147 volts 81 volts,   143 volts 81 volts,   143 volts
transfer to line power: 87 volts,   143 volts 85 volts,   139 volts 85 volts,   139 volts
battery voltage: 12 volts 36 volts 36 volts
battery recharge time: 90% of battery capacity in 2-4 hours
typical battery life: 4 to 6 years 4 to 6 years 4 to 6 years
expandable runtime: n/a n/a yes, ext battery pack
power cord: 6 ft with NEMA 5-15PHG Hospital Grade plug
receptacles: 4   (NEMA 5-15RHG Hospital Grade duplex receptacles)
communication port: DB9 USB,   DB9 USB,   DB9
surge suppression: conforms to IEEE 587 / ANSI C62.41 specs
EMI/RFI suppression: yes yes yes
phone line protection: yes n/a n/a
heat emission: 85 BTU/Hr 120 BTU/Hr 135 BTU/Hr
cooling method: convection forced air (fan) forced air (fan)
weight: 25 lbs,   11.4 kg 45 lbs,   20.5 kg 45 lbs,   20.5 kg
dimensions (in): 10.5 x 7.5 x 4.75 12.75 x 7.5 x 9 12.75 x 7.5 x 9
dimensions (cm): 26.7 x 19.1 x 12.1 32.4 x 19.1 x 22.9 32.4 x 19.1 x 22.9
operating temperature: +32º F to +104º F,     0º C to +40º C
storage temperature: +5º F to +122º F,     –15º C to +50º C
relative humidity: 0 to 95% non-condensing
Manufacturer: Tripp Lite
OEM Model: OMNISMART350HG SMART700HG SMART1200LXHG
listings: UL 1778,   UL 2601-1
FCC (emissions): Class B / part 68 Class A Class A
warranty: 2 years

features:

  • Built in isolation transformer with faraday shield
  • Limits current leakage of connected devices to below 300 uA
  • Line interactive voltage regulation corrects brownouts
  • Complete surge protection, line isolation and EMI/RFI noise filtering
  • Cold start capability allows turn-on during blackouts
  • Battery replacement door
  • Rear panel grounding lug
  • 5 diagnostic LEDs indicate:
    AC line current present
    running on battery power
    voltage correction active
    overload condition present
    battery low or needs replacement


typical applications:

  • Nurse's stations
  • Point of care solutions
  • Electronic prescribing applications
  • Computer equipment used in patient care areas
  • Telemedicine
  • Medical imaging and laboratory equipment
  • Sensitive laboratory test equipment
  • Wireless communications systems
  • OR, ICU/CCU and other patient care areas
  • Equipment with excessive leakage current
  • Equipment in areas with "dirty" power

© 2004 Supplyline