Hospital Grade UPS
specifications and product information

catalog number: 329-T045E 329-T070E 329-T140E
output capacity: 450 VA / 280 watts 700 VA / 425 watts 1400 VA / 940 watts
run time -half load: 17 minutes 22 minutes 24 minutes
run time -full load: 5 minutes 9 minutes 8 minutes
output regulation: 120 volts ±5% (while running on battery power)
frequency regulation: 60Hz ± 0.5Hz on battery power
output waveform: modified sine wave - pulse width modulated
input voltage rating: 120 volts 60Hz nominal
leakage current: less than 300 µa
maximum input current: 2.8 amps / 320 watts 3.8 amps / 460 watts 8.25 amps / 990 watts
line voltage regulation: 120 volts –18% / +8% 120 volts –18% / +8% 120 volts –18% / +8%
transfer time: 2-4 ms 2-4 ms 2-4 ms
transfer to battery: 75 volts,   147 volts 75 volts,   147 volts 75 volts,   147 volts
transfer to line power: 79 volts,   143 volts 79 volts,   143 volts 79 volts,   143 volts
battery voltage: 12 volts 12 volts 36 volts
battery recharge time: 90% of battery capacity in 2-4 hours
typical battery life: 4 to 6 years depending on use
power cord: 6 ft 14AWG with NEMA 5-15PHG Hospital Grade plug
receptacles: 4   (NEMA 5-15RHG Hospital Grade duplex receptacles)
line below 75 volts: transfer to battery power
line 75 to 92 volts: line voltage boosted by 30% at output
line 93 to 107 volts: line voltage boosted by 14% at output
line 108 to 127 volts: line voltage passed unchanged to output
line 128-147 volts: line voltage reduced by 12% at output
line above 148 volts: transfer to battery power
communications: basic DB9 monitoring port
surge suppression: conforms to IEEE 587 / ANSI C62.41 specs
EMI/RFI suppression: yes yes yes
weight: 21 lbs,   9.5 kg 22 lbs,   10 kg 37 lbs,   16.8 kg
dimensions (in): 10.75 x 5.25 x 7.25 10.75 x 5.25 x 7.25 12.5 x 7.5 x 9
dimensions (cm): 27.3 x 13.3 x 18.4 27.3 x 13.3 x 18.4 31.75 x 19.1 x 22.7
coo9ling method: convection
operating temperature: +32º F to +104º F,     0º C to +40º C
storage temperature: +5º F to +122º F,     –15º C to +50º C
relative humidity: 0 to 95% non-condensing
Manufacturer: Tripp Lite
OEM Model: OMNISMART450HG OMNISMART700HG OMNISMART1400HG
FCC (emissions): Class A Class A Class A
listings: UL 1778
warranty: 2 years

features:

  • Low leakage current and hospital grade plug connections
  • Line interactive voltage regulation corrects brownouts
  • Complete surge protection and EMI/RFI noise filtering
  • DB9 serial port for network interface; SNMP compatible
  • Cold start capability allows turn-on during blackouts
  • Battery replacement door
  • 5 diagnostic LEDs indicate:
    AC line current present
    running on battery power
    voltage correction active
    overload condition present
    battery low or needs replacement


typical applications:

  • Nurse's stations
  • Mobile medical record systems
  • Medical and dental offices
  • Laboratories
  • Medical imaging and laboratory equipment
  • Wireless communications systems
  • Where low leakage and hospital grade connections are needed